ü
Set the required temperature that is 37°C for
calibration
ü
Take 5 endotoxin test tubes (10x 75 mm)
ü
add approximately 0.5 ml of purified water in
each tube for performing the calibration
ü
Place the test tubes in the wells.
ü
about few minutes for stabilize the water within
the tubes.
ü
Place a calibrated thermometer or Temperature
RTD sensor or thermocouples in one of the tube for few seconds
ü
record the temperature. And also note down the
display temperature of heating block.
ü
After note down the reading take out and wipe
the thermometer or Temperature RTD sensor or Thermocouples with the help of a
tissue paper
ü
Continue this process for all the remaining four
tubes of LAL heating block
ü
Place the heating block on a hard vibration free
surface
ü
When incubation is in progress do not disturb
the incubator.
ACCEPTANCE CRITERIA :
37 ± 1 º C.